Users Manual Part 2
277Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
CA-0137: Generic cobas
®
infinity edge connection error
Problem
A generic cobas
®
infinity edge connection error
occurred.
Solution
Ensure that the instrument is connected to the network
and to cobas
®
infinity edge.
Check the connectivity and availability of the
cobas
®
infinity edge using the Network status test
option.
Related L1 message
I-AH
u About level 1 messages and level 2 errors (209)
CA-0138: Measuring range configuration error
Problem
The measuring range limits configured are inconsistent
and do not comply to the following rule:
Reportable range lower limit ≤ Critical range lower limit
≤ Normal range lower limit ≤ Normal range higher limit
≤ Critical range higher limit ≤ Reportable range higher
limit.
Solution
Ensure that the instrument configuration is updated with
a configuration which complies with the stated rule (via
the DMS or using the Configuration by barcode option).
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
MS
In this section
MS-0002: Measuring unit error(278)
MS-0003: Measuring unit error(278)
MS-0004: Measuring unit error(279)
MS-0005: Measuring unit error(279)
MS-0006: Measuring unit error(279)
MS-0007: Measuring unit error(280)