Users Manual Part 2

275Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Related L1 message
I-AH
u About level 1 messages and level 2 errors (209)
CA-0132: cobas
®
infinity edge execution timeout
Problem
An acknowledgement was received which exceeds the
execution acknowledgment timeout.
Solution
Check the connection between the instrument and the
cobas
®
infinity edge server using the Network status
test option.
If the problem persists, contact your Roche
representative.
Related L1 message
I-AH
u About level 1 messages and level 2 errors (209)
CA-0133: Negative cobas
®
infinity edge command acknowledgment
received
Problem
A command acknowledgement with negative outcome
code was received from cobas
®
infinity edge.
Solution
Check the connection between the instrument and the
cobas
®
infinity edge server using the Network status
test option.
If the problem persists, contact your Roche
representative.
Related L1 message
I-AH
u About level 1 messages and level 2 errors (209)
CA-0134: Service data upload failed
Problem
An error occurred while uploading service data to
cobas
®
infinity edge.
9 Error messages