CA-0130: Instrument client certificate check failed
Problem
The cobas
®
infinity edge server was not able to validate
the TLS client certificate of the instrument.
Solution
Ensure that the instrument and the cobas
®
infinity edge
server have a valid current date and time.
Check the connection between the instrument and the
cobas
®
infinity edge server using the Network status
test option.
Check your network infrastructure for breaches.
Ensure that the cobas
®
infinity edge server has the
correct TLS certificate chain installed to be able to
validate the instrument TLS (client) certificate.
If the problem persists, contact your Roche
representative.
Related L1 message
I-AH
u About level 1 messages and level 2 errors (209)
CA-0131: cobas
®
infinity edge workflow expired
Problem
A suspended cobas
®
infinity edge workflow has
exceeded the execution acknowledgment timeout.
Solution
Check the connection between the instrument and the
cobas
®
infinity edge server using the Network status
test option.
If the problem persists, contact your Roche
representative.