Users Manual Part 2

268 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
CA-0114: Password too long
Problem
The password entered is longer than the allowed 20
characters.
Solution
Enter a password which has the same or more
characters than the configured minimal password length
value, but not more than 20 characters.
Related L1 message
U-AB
u About level 1 messages and level 2 errors (209)
CA-0115: Invalid certificate configuration update
Problem
The DMS TLS certificate (DMS TLS certificate
configuration item), which was updated via the DMS, is
invalid.
The instrument restored the previous certificate.
Solution
Ensure that the DMS TLS certificate (DMS TLS
certificate configuration item) has the correct format.
If a PFX file is used, the instrument configuration must
contain the corresponding password in the DMS TLS
certificate password configuration item.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0116: Different users log on as observer before and after OTS
Problem
A user who had not logged on as observer before the
observed test sequence (OTS) tried to log on as observer
after the test.
The same observer must log in before and after the test.
Solution
Ensure that the same user is logged on as observer
before and after the observed test sequence (OTS).
9 Error messages