Users Manual Part 2

267Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Solution
Delete all lots that are no longer needed from the
instrument database (either directly on the instrument or
via the DMS).
Once an old lot is deleted, a new lot can be added to the
instrument database.
Related L1 message
G-AE
u About level 1 messages and level 2 errors (209)
CA-0111: Database limit reached for linearity lots
Problem
The database table capacity limit has been reached and
no additional lots can be saved.
Solution
Delete all lots that are no longer needed from the
instrument database (either directly on the instrument or
via the DMS).
Once an old lot is deleted, a new lot can be added to the
instrument database.
Related L1 message
G-AE
u About level 1 messages and level 2 errors (209)
CA-0113: Database limit reached for test strip lots
Problem
The database table capacity limit has been reached and
no additional lots can be saved.
Solution
Delete all lots that are no longer needed from the
instrument database (either directly on the instrument or
via the DMS).
Once an old lot is deleted, a new lot can be added to the
instrument database.
Related L1 message
G-AE
u About level 1 messages and level 2 errors (209)
9 Error messages