Users Manual Part 2
265Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Solution
Check the configuration settings in cobas
®
infinity edge
and try again with a newly generated barcode.
Related L1 message
B-AD
u About level 1 messages and level 2 errors (209)
CA-0105: Data integrity test failed
Problem
The instrument detected corrupted data in its database
and has reset the database to the default settings.
Solution
Ensure that the instrument is connected to the network
and the DMS, and that the instrument synchronizes the
configuration, patient, user, and lot data.
Related L1 message
C-AB
u About level 1 messages and level 2 errors (209)
CA-0106: Internal storage error detected
Problem
An error with the internal storage of the instrument
occurred.
Solution
Restart the instrument and try again.
If the problem persists, contact your Roche
representative.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0107: Result deletion failed
Problem
One of the result database tables is full and the
instrument was not able to delete a stored result.
9 Error messages