260 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
u About level 1 messages and level 2 errors (209)
CA-0088: Invalid patient ID
Problem
The user read a user ID via barcode that does not comply
with the maximum length allowed for the user ID.
The scanned ID is invalid.
Solution
Check if the user ID violates the maximum length
allowed of 20 characters.
Reduce the ID length to maximum 20 characters.
Related L1 message
B-AA
u About level 1 messages and level 2 errors (209)
CA-0089: Invalid patient ID
Problem
The user read an ID via RFID tag that does not comply
with the configured patterns for the patient ID.
The scanned ID is invalid.
Solution
Check if the ID complies with the intended ID patterns
(length, structure, allowed values).
Adapt the configured ID patterns (via the DMS or by
using the Configuration by barcode option) so that the
scanned barcode will be accepted by the instrument.
Related L1 message
T-AA
u About level 1 messages and level 2 errors (209)
CA-0090: Invalid patient ID
Problem
The user read an ID via RFID tag that does not comply to
the maximum allowed length for the patient ID.
The scanned ID is invalid.
9 Error messages