Users Manual Part 2

257Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0080: Entry does not exist anymore
Problem
The user tried to access a database entry or result which
has been deleted either by the instrument deletion
algorithm or by the DMS.
Solution
No action needed.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0081: Configuration by barcode error
Problem
The user scanned a barcode that does not belong to the
current barcode sequence for the configuration by
barcode (e.g., user scanned an outdated barcode from
an older barcode sequence).
Solution
Continue scanning the barcode from the correct barcode
sequence or abort the barcode sequence and restart the
scanning process.
Related L1 message
B-AC
u About level 1 messages and level 2 errors (209)
CA-0082: Low battery
Problem
No test can be performed due to low battery.
Solution
Place the instrument in the charging station in order to
charge the battery to a sufficient level.
Related L1 message
H-AC