Users Manual Part 2

255Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
CA-0074: Result database full
Problem
One of the result database tables is full and the
instrument is not able to store additional results. No
further tests can be performed.
Solution
Ensure that the instrument is connected to the network
and the DMS and that the instrument synchronizes the
configuration, patient, user, and lot data.
Check the settings of the following configuration items:
Result deletion algorithm
Result retention period (in days)
Result upload required before result deletion
Related L1 message
C-AA
u About level 1 messages and level 2 errors (209)
CA0075: Software error
Problem
The instrument is defective and cannot be used for
testing.
Solution
Contact your Roche representative.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0077: Internal storage error
Problem
An error with the internal storage of the instrument
occurred.
Solution
Restart the instrument and try again.
If the problem persists, contact your Roche
representative.
Related L1 message
G-AA
9 Error messages