Users Manual Part 2

254 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
u About level 1 messages and level 2 errors (209)
CA-0071: Invalid database encryption key
Problem
The instrument detected corrupted data in its database
and has reset the database to the default settings.
Solution
Ensure that the instrument is connected to the network
and the DMS and that the instrument synchronizes the
configuration, patient, user, and lot data.
Related L1 message
C-AB
u About level 1 messages and level 2 errors (209)
CA-0072: Database error
Problem
A database error occurred.
Solution
Contact your Roche representative.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0073: Instrument was charged during measurement
Problem
The instrument was placed into the charging station
while a measurement was in progress. The measurement
failed.
Solution
Do not place the instrument into the charging station
during a measurement.
Related L1 message
G-AC
u About level 1 messages and level 2 errors (209)
9 Error messages