Users Manual Part 2

253Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0067: Wi-Fi networks not deleted completely
Problem
The instrument could not delete all stored Wi-Fi networks
after the user tapped the Delete instrument data
button.
Solution
Restart the instrument and trigger the deletion of all
instrument data in the GUI again.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0068: Instrument restart failed
Problem
The instrument could not be restarted.
Solution
Restart the instrument.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0070: Invalid sample ID
Problem
The scanned sample ID for the proficiency material is not
valid.
Solution
Scan a valid sample ID.
The sample ID must contain between 1 and 20 printable
characters (including white space).
Related L1 message
U-AP