Users Manual Part 2

252 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Another user with the user role
"Observer" (OTS_OBSERVER) needs to log on during the
observed test sequence.
Related L1 message
U-AD
u About level 1 messages and level 2 errors (209)
CA-0063: QC / linearity lot rejected
Problem
The user tried to add a QC or linearity lot of which the
target ranges are not known to the instrument.
Solution
Ensure that the instrument receives the updated target
ranges either via barcode scanning or via DMS before
you add this QC or linearity lot.
Related L1 message
D-AN
u About level 1 messages and level 2 errors (209)
CA-0064: Measuring unit communication error
Problem
The instrument was not able to communicate with the
measuring unit.
Solution
Restart the instrument and try again.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0066: Folder not deleted
Problem
The Glucose app could not delete a folder from the file
system.
Solution
Restart the instrument and try again.
9 Error messages