Users Manual Part 2
251Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
CA-0060: Manual test strip removal
Problem
The user removed the test strip manually without using
the eject button on the user interface. This can damage
the instrument and must be avoided.
Solution
Never pull out the test strip manually.
Always us the eject button to remove the test strip.
Related L1 message
D-AO
u About level 1 messages and level 2 errors (209)
CA-0061: Insufficient user rights
Problem
The user doesn't have the privileges to perform the
requested action (e.g., being an observer for the
observed test sequence).
Solution
To allow the requested action, grant the required user
rights to the user ID in the DMS.
Ensure that the instrument is connected to the network
and the DMS synchronizes the required user rights with
the instrument.
Related L1 message
U-AQ
u About level 1 messages and level 2 errors (209)
CA-0062: Observer and logged-on user are identical
Problem
The logged-on user tried to act as observer in the
observed test sequence (OTS).
The logged-on user and the observer cannot be the
same person.
Solution
Perform the observed test sequence (OTS) with two
different users.
9 Error messages