CA-0057: Linearity lot expired
Problem
The lot of the scanned linearity material has expired.
Solution
Use a valid linearity lot.
Related L1 message
D-AE
u About level 1 messages and level 2 errors (209)
CA-0058: Unknown linearity lot scanned
Problem
The scanned linearity lot is not available on this
instrument.
Solution
Add the lot to the instrument database on the instrument
using Settings > Lot management.
Ensure that the instrument is connected to the network
in order to enable the DMS to sync the kit lot database
on the instrument.
Related L1 message
D-AN
u About level 1 messages and level 2 errors (209)
CA-0059: Multiple test strip lots used for linearity test
Problem
The user has inserted a test strip which is from a
different test strip lot than the strips used for the
previous linearity test steps.
The user must use a strip from the same test strip lot for
all steps in a linearity test.
Solution
For linearity tests, always use test strips from the same
test strip lot.
Related L1 message
D-AP
u About level 1 messages and level 2 errors (209)