CA-0053: Strip lot data request timed out
Problem
The instrument was not able to retrieve the strip lot data
from cobas infinity edge.
Solution
Ensure that the instrument is connected to the network.
Check that the instrument is connected to cobas infinity
edge using the network status test.
Related L1 message
C-AD
u About level 1 messages and level 2 errors (209)
CA-0054: Wrong kit type scanned
Problem
The user scanned a QC material bottle instead of a
linearity material bottle.
Solution
Scan the correct linearity material bottle.
Related L1 message
U-AG
u About level 1 messages and level 2 errors (209)
CA-0055: Wrong analyte scanned
Problem
The user scanned a linearity material does not match the
selected analyte.
Solution
Scan the correct linearity material that corresponds to
the analyte.
Related L1 message
U-AG
u About level 1 messages and level 2 errors (209)