248 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
CA-0050: Date / time settings lost
Problem
The current date/time settings are lost (e.g., due to loss
of power/battery change).
Solution
Enter the date and time manually and ensure that the
instrument is connected to the network.
Related L1 message
H-AF
u About level 1 messages and level 2 errors (209)
CA-0051: Invalid concentration value
Problem
The measuring unit was not able to quantify the
concentration of the measured sample.
Solution
Eject the test strip and repeat the test with a new test
strip.
If the problem persists, contact your Roche
representative.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0052: Invalid sample ID
Problem
The entered sample ID for the proficiency material is not
valid.
Solution
Enter a valid sample ID.
The sample ID must contain between 1 and 20 printable
characters (including white space).
Related L1 message
U-AO
u About level 1 messages and level 2 errors (209)