CA-0044: Measuring unit communication issue
Problem
The instrument was not able to start up the measuring
unit.
Solution
Go to the app library and open the Glucose app again. If
the problem persists, restart the instrument.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0045: Configuration update failed
Problem
The configuration barcode could not be processed.
Solution
Check the contents of the configuration barcode and
ensure that the configuration barcode is compatible with
the instrument SW version.
Related L1 message
B-AD
u About level 1 messages and level 2 errors (209)
CA-0048: Test strip lot not yet passed QC
Problem
The inserted test strip belongs to a test strip lot that has
not passed the required QC tests yet.
Solution
Required QC tests for new test strip lots is defined in the
QC algorithm settings.
Perform all required QC tests with test strips of this lot to
unlock the test strip lot for patient testing or use test strip
of a lot that already has passed all QC tests.
Related L1 message
D-AJ
u About level 1 messages and level 2 errors (209)