Users Manual Part 2
244 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
CA-0036: Wrong kit type scanned
Problem
The user scanned a linearity material bottle instead of a
QC material bottle.
Solution
Scan the correct QC material bottle.
Related L1 message
U-AG
u About level 1 messages and level 2 errors (209)
CA-0037: Wrong QC level scanned
Problem
The user scanned a QC material bottle that did not match
the selected QC level.
Solution
Scan the correct QC material level to continue the
workflow.
Related L1 message
U-AG
u About level 1 messages and level 2 errors (209)
CA-0038: QC lot expired
Problem
The lot of the scanned QC material has expired.
Solution
Use a valid QC lot.
Related L1 message
D-AE
u About level 1 messages and level 2 errors (209)
CA-0039: Unknown QC lot scanned
Problem
The scanned QC lot is not available on this instrument.
9 Error messages