242 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
u About level 1 messages and level 2 errors (209)
CA-0030: Scan timeout
Problem
The barcode scanning period has timed out. Either the
user did not scan the barcode correctly or the user tried
to scan a barcode type which is deactivated in the
instrument configuration.
Solution
Restart the barcode scanning and ensure that the
barcode is visible in the scan area on the instrument
screen.
Related L1 message
B-AB
u About level 1 messages and level 2 errors (209)
CA-0031: RFID timeout
Problem
The RFID tag reading period has timed out. Either the
user placed the RFID tag too far away from the
instrument or the user tried to scan an unsupported RFID
tag.
Solution
Restart the RFID tag reading and ensure sufficient
proximity.
Related L1 message
T-AB
u About level 1 messages and level 2 errors (209)
CA-0032: Incorrect sample applied in a QC test or linearity test
Problem
The user applied blood to the test strip instead of QC
material or linearity material.
Solution
Eject the current test strip and repeat the test with the
corresponding QC material or linearity material.
Related L1 message
S-AD
9 Error messages