CA-0027: Invalid insulin units
Problem
The user entered an value for the insulin units that
contains more than one decimal place.
Solution
Enter a valid insulin value what contains maximum 1
decimal place.
Related L1 message
U-AN
u About level 1 messages and level 2 errors (209)
CA-0028: Invalid user ID
Problem
The user entered a user ID that does not comply to the
configured pattern for the user ID.
Solution
Check if the entered user ID which was rejected by the
instrument complies to the intended user ID pattern
(length, structure, allowed values).
Adapt the configured pattern if the entered user ID shall
be accepted by the instrument.
Related L1 message
U-AA
u About level 1 messages and level 2 errors (209)
CA-0029: User ID not found
Problem
The scanned user ID is unknown to the instrument.
Solution
Ensure that the instrument is connected to the network.
Check the connectivity and availability of the DMS using
the network status test.
Ensure that the DMS synchronizes the user list with the
instrument.
Related L1 message
U-AJ