Users Manual Part 2

238 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
CA-0017: QC lot expired
Problem
The QC lot used has expired.
Solution
Repeat the test using a valid QC lot.
Related L1 message
D-AE
u About level 1 messages and level 2 errors (209)
CA-0019: User ID not found
Problem
The scanned user ID is unknown to the instrument.
Solution
Ensure that the instrument is connected to the network
and the DMS. Check the connectivity and availability of
the DMS using the network status test.
Ensure that the DMS synchronizes the user list with the
instrument.
Related L1 message
U-AJ
u About level 1 messages and level 2 errors (209)
CA-0020: Patient ID not found
Problem
The scanned patient ID is unknown to the instrument.
Solution
Ensure that the instrument is connected to the network
and the DMS. Check the connectivity and availability of
the DMS using the network status test.
Ensure that the DMS synchronizes the patient list with
the instrument.
Related L1 message
U-AK
u About level 1 messages and level 2 errors (209)