Users Manual Part 2
236 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
u About level 1 messages and level 2 errors (209)
CA-0011: Software error
Problem
The instrument is defective and cannot be used for
testing.
Solution
Contact your Roche representative.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0012: Synchronization with DMS failed
Problem
The synchronization with the DMS failed. The instrument
cannot transmit results or receive updated data from the
DMS.
Solution
Ensure that the instrument is connected to the network
and the DMS. Check the connectivity and availability of
the DMS using the network status test.
Related L1 message
I-AG
u About level 1 messages and level 2 errors (209)
CA-0013: Test strip inserted too early
Problem
The test strip was inserted too early.
Solution
Eject the test strip and insert it again when requested.
Related L1 message
D-AK
u About level 1 messages and level 2 errors (209)
9 Error messages