Users Manual Part 2

235Error messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
CA-0008: Test strip lot information not found
Problem
The lot information of the inserted test strip is missing.
The inserted test strip cannot be used
Solution
Add the unknown test strip lot to the instrument
database by using one of the following methods:
Scan the strip vial container and let the instrument
request the lot from cobas
®
infinity edge.
Scan the complete lot information via a configuration
barcode (generated in the cobas
®
infinity edge
management portal).
If the lot is already known in the DMS, configure the
DMS to upload the lot information to the instrument.
Alternatively the user could use a test strip from a
known/valid test strip lot.
Related L1 message
D-AH
u About level 1 messages and level 2 errors (209)
CA-0009: Test strip lot expired
Problem
The used test strip lot is expired.
Solution
Repeat the test with a test strip from a valid lot.
Related L1 message
D-AB
u About level 1 messages and level 2 errors (209)
CA-0010: Communication with measuring unit interrupted
Problem
The communication with the measuring unit was
interrupted.
Solution
Restart the instrument and try again.
Related L1 message
G-AA
9 Error messages