Users Manual Part 2
234 Level 2 errors
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
u About level 1 messages and level 2 errors (209)
CA-0005: Measuring unit communication error
Problem
The instrument was not able to communicate with the
measuring unit.
Solution
Restart the instrument and try again.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0006: Access to internal storage failed
Problem
The instrument was not able to access its internal
storage.
Solution
Restart the instrument and try again.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
CA-0007: Software error
Problem
The instrument is defective and cannot be used for
testing.
Solution
Contact your Roche representative.
Related L1 message
G-AA
u About level 1 messages and level 2 errors (209)
9 Error messages