Users Manual Part 2
480 Generic instrument functionality
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Cleaning notification after each patient test(480)
Time interval for cleaning notification (in hours)
This configuration item specifies the time in hours since
instrument cleaning was last performed and when a
cleaning notification is displayed.
The notification is displayed on the Settings >
Notifications screen. You confirm that the instrument
has been cleaned by choosing the Confirm button on
the Instrument cleaning dialog box.
The time interval to the next cleaning alert commences
with the confirmation.
Item ID
CLEANING_ALERT_HOURS
Available values/format
• INT:0-999 hours
• Default value = "0" hours
Configuration
You configure the Time interval for cleaning
notification (in hours) configuration item using the
following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
u Related topics
• List of notifications (72)
Cleaning notification after each patient test
This configuration item determines if the user is notified
to clean the instrument after a patient test has been
made.
The notification is displayed on the Settings >
Notifications screen.
Item ID
CLEANING_REQUIRED
Available values/format
• Disabled (default)
[false]
A cleaning notification is not displayed after a patient
test has been made.
• Enabled
[true]
11 General configuration