477General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Available values/format
• Disabled (default)
[false]
Patient test series disabled.
• Enabled
[true]
Patient test series enabled.
Configuration
You configure the Patient test series configuration item
using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
Allow linearity tests
This configuration item specifies if linearity tests can be
performed on the instrument.
When the configuration item is enabled the Linearity
button is available on the Tests screen. When it is
disabled the button is excluded.
Item ID
LINEARITY_ENABLED
Available values/format
• Enabled (default)
[true]
Linearity tests can be performed.
• Disabled
[false]
Linearity tests cannot be performed.
Configuration
You configure the Allow linearity tests configuration
item using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
Allow proficiency tests
This configuration item specifies if proficiency tests can
be performed on the instrument.
When the configuration item is enabled the Proficiency
button is available on the Tests screen. When it is
disabled the button is excluded.
Item ID
PROFICIENCY_ENABLED
11 General configuration