RFID tag configuration
This configuration item specifies what is used for
identification when using a RFID tag.
Item ID
RFID_TAG_CONFIGURATION
Available options/values
Unique RFID tag ID (default)
[UNIQUE_ID]
The unique ID of the tag is used (between 4 bytes to
up to 11 bytes).
Stored NDEF record
[NDEF_RECORD]
An NDEF record stored on the tag is used (the record
number is configured using the User RFID tag
record configuration configuration item).
Dependencies
The User authentication required configuration item
has a value of RFID.
Configuration
You configure the RFID tag configuration configuration
item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
Tests menu
In this section
Patient test series(476)
Allow linearity tests(477)
Allow proficiency tests(477)
Insulin dose entry(478)
Startup behavior of the instrument(478)
Message of the day(479)
Patient test series
This configuration item specifies if a patient test series
can be performed (i.e., performing successive tests on
the same patient).