Users Manual Part 2
476 Generic instrument functionality
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
RFID tag configuration
This configuration item specifies what is used for
identification when using a RFID tag.
Item ID
RFID_TAG_CONFIGURATION
Available options/values
• Unique RFID tag ID (default)
[UNIQUE_ID]
The unique ID of the tag is used (between 4 bytes to
up to 11 bytes).
• Stored NDEF record
[NDEF_RECORD]
An NDEF record stored on the tag is used (the record
number is configured using the User RFID tag
record configuration configuration item).
Dependencies
The User authentication required configuration item
has a value of RFID.
Configuration
You configure the RFID tag configuration configuration
item using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
Tests menu
In this section
Patient test series(476)
Allow linearity tests(477)
Allow proficiency tests(477)
Insulin dose entry(478)
Startup behavior of the instrument(478)
Message of the day(479)
Patient test series
This configuration item specifies if a patient test series
can be performed (i.e., performing successive tests on
the same patient).
Item ID
PATIENT_TEST_SERIES
11 General configuration