Users Manual Part 2

472 Generic instrument functionality
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Available options/values
Enabled (default)
[true]
List selection activated.
A linearity lot number can be selected from a list or
by barcode scan.
Disabled
[false]
List selection deactivated.
A linearity lot number can be selected by barcode
scan only.
Dependencies
The Allow linearity lot selection from list
configuration item is only relevant when the Allow
linearity tests configuration item has a value of
Enabled.
Configuration
You configure the Allow linearity lot selection from
list configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
Allow QC lot selection from list
This configuration item determines how a user selects a
QC lot number. Either from a list or by barcode scan, or
by barcode only.
The list can be activated and deactivated.
Item ID
QC_LOT_SELECTION_BY_LIST
Available options/values
Enabled (default)
[true]
List selection activated.
A QC lot number can be selected from a list or by
barcode scan.
Disabled
[false]
List selection deactivated.
A QC lot number can be selected by barcode scan
only.
Configuration
You configure the Allow QC lot selection from list
configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
11 General configuration