Users Manual Part 2
471General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Allow access to lot management on the instrument
This configuration item determines access to the
Settings > Lot management option.
Item ID
LOT_MANAGEMENT_ACCESS
Available options/values
• Disabled
[DISABLED]
Access to Settings > Lot management option is
disabled.
Lot management is only possible using
cobas
®
infinity edge and/or the DMS.
• Enabled (no restrictions)
[UNLIMITED_ACCESS]
Anyone can access the Settings > Lot
management option.
• Enabled (setup password required) (default)
[VIA_SETUP_PASSWORD]
The Instrument settings password is required to
access the Settings > Lot management option.
• Enabled (corresponding user rights required)
[VIA_OPERATOR_PRIVILEGE]
Access to the Settings > Lot management option
is only available to users with the "POCC" privilege.
Configuration
You configure the Allow access to lot management
on the instrument configuration item using the
following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
Lot management
In this section
Allow linearity lot selection from list(471)
Allow QC lot selection from list(472)
Allow linearity lot selection from list
This configuration item determines how a user selects a
linearity lot number. Either from a list or by barcode scan,
or by barcode scan only.
Item ID
LINEARITY_LOT_SELECTION_BY_LIST
11 General configuration