Users Manual Part 2

469General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
The instrument settings password must be entered in
order to access the Settings > Configuration by
barcode option.
Enabled (corresponding user rights required)
[VIA_OPERATOR_PRIVILEGE]
Access to the Settings > Configuration by
barcode option is only available to users with the
required ("POCC") privilege.
Configuration
You configure the Allow configuration by barcode on
the instrument configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
Instrument settings password (467)
Access restrictions to date and time
This configuration item determines access to the
Settings > Date and time option in which you set the
instrument date and time.
Item ID
DATE_TIME_EDITING_ACCESS
Available options/values
Only via DMS
[VIA_DMS]
Setting the instrument date and time is only possible
via DMS/POCT1-A.
No restrictions (default)
[UNLIMITED_ACCESS]
The instrument date and time can be set by anyone.
Setup password required
VIA_SETUP_PASSWORD
The Instrument settings password is required for
setting the instrument date and time.
Corresponding user rights required
[VIA_OPERATOR_PRIVILEGE]
The instrument date and time can only be set by a
user with the required ("POCC") privilege.
Dependencies
The Access restrictions to date and time
configuration item is only active when the Network
Time Protocol (NTP) configuration item has a value of
Disabled.
11 General configuration