Users Manual Part 2

460 Measurement workflow
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Barcode (generated via cobas
®
infinity edge)
Sample type
This configuration item specifies the available sample
types that can be displayed to the user when performing
a glucose test.
The Sample type configuration item is used to indicate
the source of the blood sample. If it has no value
(default), the sample type selection screen is not
displayed.
When the configuration item contains values, the sample
type selection screen is displayed listing the
corresponding sample types.
Item ID
SAMPLE_TYPE_SELECTION
Available values/format
Capillary blood
[CAPILLARY]
Venous blood
[VENOUS]
Arterial blood
[ARTERIAL]
Heel-stick blood
[HEEL_STICK]
"No value" (default)
Configuration
You configure the Sample type configuration item using
the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
Result unit (glucose tests)
This configuration item specifies measurement unit used
in glucose test results.
If the result unit is changed, the instrument deletes all
stored results and resets all configured measurement
range values to their default values.
You must ensure that all unsent results are retrieved
from the instrument before changing the value of this
configuration item.
11 General configuration