Users Manual Part 2
459General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
• Barcode (generated via cobas
®
infinity edge)
u Related topics
• Default patient ID for emergency tests (459)
Default patient ID for emergency tests
This configuration item specifies the ID that is used as
the patient ID for emergency glucose tests.
Item ID
EMERGENCY_ID
Available values/format
• 1-20 alphanumeric characters
• "99999999" (default)
Dependencies
The Allow emergency tests configuration item has the
value Enabled.
Configuration
You configure the Default patient ID for emergency
tests configuration item using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
u Related topics
• Allow emergency tests (458)
Test strip ejection
This configuration item specifies at which point in the
glucose test workflow the user ejects a used test strip.
Item ID
STRIP_EJECT_CONFIGURATION
Available values/format
• After test result confirmation (default)
[AFTER_RESULT_CONFIRMATION]
Test strip is ejected after the test result confirmation.
• Before test result confirmation
[BEFORE_RESULT_CONFIRMATION]
Test strip is ejected before the test result
confirmation.
Configuration
You configure the Test strip ejection configuration item
using the following:
• Data management system (e.g., cobas
®
infinity POC)
11 General configuration