Users Manual Part 2
457General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Configuration
You configure the Message displayed if the result is
above the critical range configuration item using the
following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
u Related topics
• Critical range upper limit (450)
• Custom message (above critical range) (455)
Message displayed if the result is below the critical range
This configuration item determines if the "below critical
range message" is displayed when a glucose test result
value is below the critical range.
The content of the custom message is entered using the
Custom message (below critical range)
configuration item.
If a custom message has not been configured, a built-in
message is displayed when the value for this
configuration item is Enabled.
Item ID
DISPLAY_CRITICAL_RANGE_LO_MESSAGE
Available values/format
• Disabled (default)
[false]
The message is not displayed.
• Enabled
[true]
The message is displayed.
Configuration
You configure the Message displayed if the result is
below the critical range configuration item using the
following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
u Related topics
• Custom message (below critical range) (456)
• Critical range lower limit (449)
11 General configuration