Users Manual Part 2

457General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Configuration
You configure the Message displayed if the result is
above the critical range configuration item using the
following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
Critical range upper limit (450)
Custom message (above critical range) (455)
Message displayed if the result is below the critical range
This configuration item determines if the "below critical
range message" is displayed when a glucose test result
value is below the critical range.
The content of the custom message is entered using the
Custom message (below critical range)
configuration item.
If a custom message has not been configured, a built-in
message is displayed when the value for this
configuration item is Enabled.
Item ID
DISPLAY_CRITICAL_RANGE_LO_MESSAGE
Available values/format
Disabled (default)
[false]
The message is not displayed.
Enabled
[true]
The message is displayed.
Configuration
You configure the Message displayed if the result is
below the critical range configuration item using the
following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
Custom message (below critical range) (456)
Critical range lower limit (449)
11 General configuration