Users Manual Part 2

454 Result range handling
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Message displayed if the result is above the reportable
range
This configuration item determines if the "above
reportable range message" is displayed when a glucose
test result value is above the reportable range.
The content of the custom message is entered using the
Custom message (above reportable range)
configuration item.
If a custom message has not been configured, a built-in
message is displayed when the value for this
configuration item is Enabled.
Item ID
DISPLAY_REPORTABLE_RANGE_HI_MESSAGE
Available values/format
Disabled (default)
[false]
The message is not displayed.
Enabled
[true]
The message is displayed.
Configuration
You configure the Message displayed if the result is
above the reportable range configuration item using
the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
Custom message (above reportable range) (454)
Reportable range upper limit (451)
Custom message (above reportable range)
You use this configuration item to create the content of
the above reportable range message.
The message is displayed when the following conditions
are true:
The glucose test result is greater than the value set in
the Reportable range upper limit configuration
item.
11 General configuration