Users Manual Part 2

450 Result range handling
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Item ID
CRITICAL_RANGE_LO_LIMIT
Available values/format
Default values:
10 mg/dL
0.6 mmol/L
Dependencies
The following dependencies must be true:
The Critical range lower limit value must be greater
than or equal to the Reportable range lower limit
value.
The Critical range lower limit value must be less
than or equal to the Normal range lower limit
value.
Configuration
You configure the Critical range lower limit
configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
Message displayed if the result is below the critical
range (457)
Critical range upper limit
This configuration item specifies the value of the upper
limit of the critical range for glucose test results. It is
expressed in mg/dL or mmol/L.
The default value is set by the Glucose app based on the
Result unit (glucose tests) configuration item.
Item ID
CRITICAL_RANGE_HI_LIMIT
Available values/format
Default values:
600 mg/dL
33.3 mmol/L
Dependencies
The following dependencies must be true:
The Critical range upper limit value must be
greater than or equal to the Normal range upper
limit value.
The Critical range upper limit value must be less
than or equal to the Reportable range upper limit
value.
11 General configuration