Users Manual Part 2

449General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
The Normal range lower limit value must be
greater than or equal to the Critical range lower
limit value.
Configuration
You configure the Normal range lower limit
configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
Normal range upper limit
This configuration item specifies the value of the upper
limit of the normal range for glucose test results. It is
expressed in mg/dL or mmol/L.
The default value is set by the Glucose app based on the
Result unit (glucose tests) configuration item.
Item ID
NORMAL_RANGE_HI_LIMIT
Available values/format
Default values:
600 mg/dL
33.3 mmol/L
Dependencies
The following dependencies must be true:
The Normal range upper limit value must be
greater than the Normal range lower limit value.
The Normal range upper limit value must be less
than or equal to the Critical range upper limit
value.
Configuration
You configure the Normal range upper limit
configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
Critical range lower limit
This configuration item specifies the value of the lower
limit of the critical range for glucose test results. It is
expressed in mg/dL or mmol/L.
The default value is set by the Glucose app based on the
Result unit (glucose tests) configuration item.
11 General configuration