Users Manual Part 2
448 Result range handling
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Result range handling
In this section
Normal range lower limit(448)
Normal range upper limit(449)
Critical range lower limit(449)
Critical range upper limit(450)
Reportable range lower limit(451)
Reportable range upper limit(451)
Message displayed if the result is below the reportable
range(452)
Custom message (below reportable range)(453)
Message displayed if the result is above the reportable
range(454)
Custom message (above reportable range)(454)
Custom message (above critical range)(455)
Custom message (below critical range)(456)
Message displayed if the result is above the critical
range(456)
Message displayed if the result is below the critical
range(457)
Normal range lower limit
This configuration item specifies the value of the lower
limit of the normal range for glucose test results. It is
expressed in mg/dL or mmol/L.
The default value is set by the Glucose app based on the
Result unit (glucose tests) configuration item.
Item ID
NORMAL_RANGE_LO_LIMIT
Available values/format
Default values:
• 10 mg/dL
• 0.6 mmol/L
Dependencies
The following dependencies must be true:
• The Normal range lower limit value must be less
than the Normal range upper limit value.
11 General configuration