Users Manual Part 2

448 Result range handling
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Result range handling
In this section
Normal range lower limit(448)
Normal range upper limit(449)
Critical range lower limit(449)
Critical range upper limit(450)
Reportable range lower limit(451)
Reportable range upper limit(451)
Message displayed if the result is below the reportable
range(452)
Custom message (below reportable range)(453)
Message displayed if the result is above the reportable
range(454)
Custom message (above reportable range)(454)
Custom message (above critical range)(455)
Custom message (below critical range)(456)
Message displayed if the result is above the critical
range(456)
Message displayed if the result is below the critical
range(457)
Normal range lower limit
This configuration item specifies the value of the lower
limit of the normal range for glucose test results. It is
expressed in mg/dL or mmol/L.
The default value is set by the Glucose app based on the
Result unit (glucose tests) configuration item.
Item ID
NORMAL_RANGE_LO_LIMIT
Available values/format
Default values:
10 mg/dL
0.6 mmol/L
Dependencies
The following dependencies must be true:
The Normal range lower limit value must be less
than the Normal range upper limit value.
11 General configuration