Users Manual Part 2

444 User ID handling
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Item ID
OPERATOR_ID_PATTERN_RFID
Available values/format
ST:0-500 (regular expression)
No default value set.
Configuration
You configure the User ID pattern validation (RFID)
configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
User ID validation (439)
User authentication type (438)
User ID handling
In this section
Minimum password length(444)
Logon session timeout(445)
Show user lockout warning prior to user certificate
expiration(445)
User ID autocomplete(446)
Minimum password length
This configuration item specifies the minimum length
(characters) of a new password which is validated by the
instrument when entered by a user.
Item ID
MINIMAL_PASSWORD_LENGTH
Available values/format
4 - 20 characters
Default = 4 characters
Dependencies
The Minimum password length configuration item is
only relevant when the following 2 conditions are true:
The User authentication required configuration
item has a value of Enabled or Enabled for QC test
only.