Users Manual Part 2
429General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
In this section
Patient ID validation(429)
Patient ID pattern validation (keyboard)(430)
Patient ID pattern validation (barcode)(431)
Patient ID pattern validation (RFID)(433)
Patient ID validation
This configuration item specifies how a patient ID is
validated when entered.
Item ID
PATIENT_ID_VALIDATION
Available values/format
• Pattern-based (default)
[PATTERN_BASED_ONLY]
The patient ID entered is validated against the pattern
defined by the corresponding configuration item.
• Patient ID pattern validation (keyboard)
• Patient ID pattern validation (barcode)
• Patient ID pattern validation (RFID)
• List-based
[LIST_BASED]
The patient ID entered is validated in the same way as
with the Pattern-based value. It is also validated
against a patient list which has been provided by the
DMS.
• List-based (allow unknown patient IDs)
[LIST_BASED_ALLOW_NEW_ENTRIES_IF_NOT_IN_LI
ST]
The patient ID entered is validated in the same way as
with the Pattern-based value. It is also validated
against a patient list which has been provided by the
DMS.However, if the patient ID is not found in the
instrument’s patient list, and it is successfully
validated against the pattern defined by the
corresponding configuration item, the user may still
run the test for the patient and record the result.
Dependencies
For cobas
®
infinity edge only, Pattern-based is the is
the only valid value.
For cobas
®
infinity edge and a DMS, if the DMS
supports patient lists, List-based and List-based
(allow unknown patient IDs) are valid values.
11 General configuration