Patient ID handling
In this section
Patient ID entry interfaces(427)
Patient ID validation(428)
Patient ID confirmation dialog(434)
Patient ID autocomplete(435)
Allowed barcode types for patient ID(436)
Patient ID entry interfaces
In this section
Patient entry type(427)
Patient entry keyboard type(428)
Patient entry type
This configuration item specifies the available interfaces
with which a patient ID is entered. Combinations of
interfaces can be configured (e.g., Barcode and
Keyboard, or Keyboard and RFID).
i
When the Patient entry type configuration item
has no value the configuration is rejected.
Item ID
PATIENT_ID_ENTRY_TYPE
Available values/format
Barcode (default)
[BARCODE]
Patient ID is entered using a barcode scan.
Keyboard (default)
[KEYBOARD]
Patient ID is entered using the virtual keyboard.
RFID
[RFID]
Patient ID is entered using an RFID tag.
Dependencies
Barcode
Patient ID pattern validation (barcode)
Allowed barcode types for patient ID
Checksum validation of barcode type
"Interleaved 2 of 5"
11 General configuration