Users Manual Part 2
425General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
List of power management options
In this section
Standby timeout(425)
Lift to wake(425)
Standby timeout
This configuration item specifies the time in seconds that
an instrument is idle before switching to standby mode.
Item ID
ACTIVITY_TURN_OFF_IN_SECONDS
Available values/format
30-3600 seconds
(default=300)
Configuration
You configure the Standby timeout configuration item
using the following methods:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
Lift to wake
This configuration item specifies whether or not the
instrument wakes from standby when lifted from rest
(movement is detected).
Item ID
LIFT_TO_WAKE
Available values/format
• Disabled (default)
[false]
Movement detection is disabled.
• Enabled
[true]
Movement detection is enabled and the instrument
wakes up when lifted.
Configuration
You configure the Lift to wake configuration item using
the following methods:
11 General configuration