Users Manual Part 2

425General configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
List of power management options
In this section
Standby timeout(425)
Lift to wake(425)
Standby timeout
This configuration item specifies the time in seconds that
an instrument is idle before switching to standby mode.
Item ID
ACTIVITY_TURN_OFF_IN_SECONDS
Available values/format
30-3600 seconds
(default=300)
Configuration
You configure the Standby timeout configuration item
using the following methods:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
Lift to wake
This configuration item specifies whether or not the
instrument wakes from standby when lifted from rest
(movement is detected).
Item ID
LIFT_TO_WAKE
Available values/format
Disabled (default)
[false]
Movement detection is disabled.
Enabled
[true]
Movement detection is enabled and the instrument
wakes up when lifted.
Configuration
You configure the Lift to wake configuration item using
the following methods:
11 General configuration