Users Manual Part 2
402 Connectivity
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Available values/format
0-60 characters
For a host specified in literal IPv6 address, only the
“preferred form” according to RFC 2373 (paragraph 2.2
1.) is supported.
Compressing zeros, mixed IPv6/IPv4 representations or
scoped addresses are not allowed.
Examples:
• fd00:0:0:0:4533:4431:55b5:b46 is supported.
• fd00:0000:0000:0000:4533:4431:55b5:b46 is
supported.
• fd00::4533:4431:55b5:b46 is not supported due to
compressed zeros.
• fd00:0000:0000:0000:4533:4431:85.181.11.70 is not
supported due to mixed IPv6/IPv4 notation.
Adherence to the permitted form can be checked
using the following regular expression:
((?:[A-Fa-f0-9]{1,4}:){7}[A-Fa-f0-9]{1,4})?
Dependencies
cobas
®
infinity edge only:
If the instrument will only be connected to
cobas
®
infinity edge, the cobas infinity edge IP
address / host name and cobas infinity edge port
number configuration items must be set.
cobas
®
infinity edge and DMS:
The cobas infinity edge IP address / host name
configuration item must be set by the DMS.
Configuration
You configure the cobas infinity edge IP address /
host name configuration item using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
cobas infinity edge port number
This configuration item specifies the "cobas infinity edge
port number".
Item ID
RSG_PORT
Available values/format
0 (default)-65535 characters
10 Initial configuration