Users Manual Part 2

400 Connectivity
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Data management system IP address / host name
This configuration item specifies the DMS IP address /
host name.
It can either be a machine name, such as
"dms.hospital.com", or a textual representation of its IP
address. If a literal IP address is supplied, only the validity
of the address format is checked.
Item ID
DMS_Host
Available values/format
0-60 characters
For a host specified as a literal IPv6 address, only the
“preferred form” according to RFC 2373 (paragraph 2.2
1.) is supported.
Compressing zeros, mixed IPv6/IPv4 representations or
scoped addresses are not supported.
Examples:
fd00:0:0:0:4533:4431:55b5:b46 is supported.
fd00:0000:0000:0000:4533:4431:55b5:b46 is
supported.
fd00::4533:4431:55b5:b46 is not supported due to
compressed zeros.
fd00:0000:0000:0000:4533:4431:85.181.11.70 is not
supported due to mixed IPv6/IPv4 notation.
Adherence to the permitted form can be checked using
the following regular expression: ((?:[A-Fa-f0-9]{1,4}:){7}
[A-Fa-f0-9]{1,4})?
If this configuration item remains empty, the instrument
will not communicate to any DMS and therefore will not
log any issues related to DMS communication in the
audit trail.
Dependencies
cobas
®
infinity edge only:
If the instrument will only be connected to
cobas
®
infinity edge, the Data management system IP
address / host name configuration item must not be set.
cobas
®
infinity edge and DMS:
The Data management system IP address / host
name and the Data management system port number
must be set.
10 Initial configuration