Users Manual Part 2
399Initial configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Dependencies
Depending on the DHCP server capabilities, the
instrument automatically receives an NTP server address.
If the DHCP server does not provide an NTP server
address, the NTP server address configuration item
must be set to a valid value.
Configuration
You configure the NTP server address configuration
item using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
Instrument time zone
This configuration item determines the time zone of the
instrument taken from a subset of the ICANN/IANA time
zone database (also referred to as Olson database).
This time zone is used as the instrument time zone and
will be taken into account when the instrument receives
the actual time via NTP or via POCT1-A directive.
If the Instrument time zone configuration item is not
set to a dedicated time zone (e.g., Europe/Berlin), then
the daylight saving logic is not active.
Item ID
TIME_ZONE
Available values/format
• 0-50 characters
• Etc./GMT (default)
• For a list of valid time zones refer to the appropriate
subset of the ICANN/IANA time zone database.
Dependencies
The Network Time Protocol (NTP) configuration item
is set to Enabled.
Configuration
You configure the Instrument time zone configuration
item using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
10 Initial configuration