Users Manual Part 2

397Initial configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Adherence to the supported form can be checked using
the following regular expression:
((?:[A-Fa-f0-9]{1,4}:){7}[A-Fa-f0-9]{1,4})?
Dependencies
If this configuration item is set, the Instrument IPv4
address configuration item must not be used.
If the Instrument IPv4 address configuration item is
set, the instrument will ignore the value set in this
configuration item as the operating system only allows
one static IP address to be set.
Configuration
You configure the Instrument IPv6 address
configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
IPv4 subnet mask
This configuration item specifies the IPv4 subnet mask.
Item ID
IP4_SUBNET_MASK
Available values/format
0.0.0.0 – 255.255.255.255
Dependencies
If the Use DHCP configuration item has a value of
Yes, the instrument will normally receive this value
automatically (depending on the DHCP server
configuration) and this configuration item can remain
empty.
If the Use DHCP configuration item has a value of
No, this configuration item must be set.
Configuration
You configure the IPv4 subnet mask configuration item
using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
Network Time Protocol (NTP)
This configuration item specifies if the instrument uses
the Network Time Protocol (NTP).
10 Initial configuration