Users Manual Part 2
396 Connectivity
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Item ID
INSTRUMENT_IP4_ADDRESS
Available values/format
0.0.0.0 – 255.255.255.255
Dependencies
• If the Use DHCP configuration item has a value of
Yes, the instrument will normally receive this value
automatically (depending on the DHCP server
configuration) and this configuration item can remain
empty.
• If the Use DHCP configuration item has a value of
No, this configuration item must be set.
• If this configuration item is set, the instrument will
ignore the value set in the Instrument IPv6 address
DNS IPv6 address configuration item as the
operating system only allows one static IP address to
be set.
Configuration
You configure the Instrument IPv4 address
configuration item using the following:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
Instrument IPv6 address
This configuration item specifies the instrument IPv6
address with IPv6 prefix.
Item ID
INSTRUMENT_IP6_ADDRESS
Available values/format
Only the “preferred form”
(6)
is supported. Compressing
zeros, mixed IPv6/IPv4 representations or scoped
addresses are not supported.
Examples:
fd00:0:0:0:4533:4431:55b5:b46 is supported.
fd00:0000:0000:0000:4533:4431:55b5:b46 is supported.
fd00::4533:4431:55b5:b46 is not supported due to
compressed zeros.
fd00:0000:0000:0000:4533:4431:85.181.11.70 is not
supported due to mixed IPv6/IPv4 notation.
(6)
RFC 2373 (paragraph 2.2 1.)