Users Manual Part 2

391Initial configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
If a value of "0" is selected no warning is given.
Item ID
DMS_SYNCHRONIZATION_LOCKOUT_HOURS
Available values/format
0-999 hours
Default = "0"
Configuration
You configure the Show synchronization lockout
warning prior to actual lockout (in hours)
configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
Synchronization lockout (in hours) (391)
Synchronization lockout (in hours)
This configuration item specifies the time in hours
between a data synchronization warning and a
synchronization lockout if no synchronization takes
place.
It works in conjunction with the Show synchronization
lockout warning prior to actual lockout (in hours)
configuration item.
If a value of "0" is selected synchronization lockout is
disabled.
Item ID
DMS_SYNCHRONIZATION_LOCKOUT_HOURS
Available values/format
0-999 hours
Default = "0"
Configuration
You configure the Synchronization lockout (in hours)
configuration item using the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
Show synchronization lockout warning prior to actual
lockout (in hours) (390)
10 Initial configuration