Users Manual Part 2

384 QC configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
The custom deviation configuration depends upon the
QC version. The QC version defines the target values
(level 1 and level 2) of a QC kit provided by Roche. The
QC version ranges from 0 to 9.
The QC version of a QC kit is located in Settings > Lot
management > Lot management > QC lots. The
initial QC version at product launch is 0.
Item ID
GLU_QC_CUSTOM_DEVIATION_L2
Available values/format
SET of value pairs (comma separated):
The accepted maximum deviation values (for result
unit mg/dL or mmol/L) are defined by Roche (see
package insert of cobas
®
GLU QC kit). Custom
deviation values which exceed the permitted
deviation defined by Roche will be rejected by the
instrument.
If the configuration item is empty (default) or the QC
version is set to "0:0", no custom deviation is used.
If a QC version is listed twice (misconfiguration), the
value of the last entry in the list is used as the custom
deviation.
Example 1:
"0:35" = using a custom deviation of 35 mg/dL for QC
version 0.
Example 2:
"0:1.9" = using a custom deviation of 1.9 mmol/L for QC
version 0.
Example3:
"0:35,1:37" = using a custom deviation of 35 mg/dL for
QC version 0 and 37 mg/dL for QC version 1.
Example4:
"0:0" = no custom deviation is used.
Dependencies
The allowed deviation is dependent on the Result unit
(glucose tests) configuration item, i.e. mmol/L or mg/
dL.
Configuration
You configure the Allowed deviation for QC level 2
result configuration item using the following methods:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
10 Initial configuration