• The accepted maximum deviation values (for result
unit mg/dL or mmol/L) are defined by Roche (see
package insert of cobas
®
GLU QC kit). Custom
deviation values which exceed the permitted
deviation defined by Roche will be rejected by the
instrument.
• If the configuration item is empty (default) or the QC
version is set to "0:0", no custom deviation is used.
• If a QC version is listed twice (misconfiguration), the
value of the last entry in the list is used as the custom
deviation.
Example 1:
"0:6" = using a custom deviation of 6 mg/dL for QC
version 0.
Example 2:
"0:0.3" = using a custom deviation of 0.3 mmol/L for QC
version 0.
Example3:
"0:6,1:7" = using a custom deviation of 6 mg/dL for QC
version 0 and 7mg/dL for QC version 1.
Example4:
"0:0" = no custom deviation is used.
Dependencies
The allowed deviation is dependent on the Result unit
(glucose tests) configuration item, i.e. mmol/L or mg/
dL.
The instrument rejects custom deviation values which
exceed the allowed deviation defined by Roche. See the
package insert of cobas
®
GLU QC kit for the deviations
defined by Roche.
Configuration
You configure the Allowed deviation for QC level 1
result configuration item using the following methods:
• Data management system (e.g., cobas
®
infinity POC)
• Barcode (generated via cobas
®
infinity edge)
Allowed deviation for QC level 2 result
This configuration item specifies the permissible
deviation from the level 2 target value for which a QC
result is valid.