Users Manual Part 2

383Initial configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
The accepted maximum deviation values (for result
unit mg/dL or mmol/L) are defined by Roche (see
package insert of cobas
®
GLU QC kit). Custom
deviation values which exceed the permitted
deviation defined by Roche will be rejected by the
instrument.
If the configuration item is empty (default) or the QC
version is set to "0:0", no custom deviation is used.
If a QC version is listed twice (misconfiguration), the
value of the last entry in the list is used as the custom
deviation.
Example 1:
"0:6" = using a custom deviation of 6 mg/dL for QC
version 0.
Example 2:
"0:0.3" = using a custom deviation of 0.3 mmol/L for QC
version 0.
Example3:
"0:6,1:7" = using a custom deviation of 6 mg/dL for QC
version 0 and 7mg/dL for QC version 1.
Example4:
"0:0" = no custom deviation is used.
Dependencies
The allowed deviation is dependent on the Result unit
(glucose tests) configuration item, i.e. mmol/L or mg/
dL.
The instrument rejects custom deviation values which
exceed the allowed deviation defined by Roche. See the
package insert of cobas
®
GLU QC kit for the deviations
defined by Roche.
Configuration
You configure the Allowed deviation for QC level 1
result configuration item using the following methods:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
Allowed deviation for QC level 2 result
This configuration item specifies the permissible
deviation from the level 2 target value for which a QC
result is valid.
10 Initial configuration