Users Manual Part 2

381Initial configuration
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
"06:00,12:00,18:00"
"02:00,06:00,10:00,14:00,18:00,22:00"
"06:00" (default)
Dependencies
Configuring Time of day is only relevant when the QC
algorithm configuration item value is Time of day or
Time of day (alternating QC level).
Configuration
You configure the Time of day configuration item using
the following:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
u Related topics
QC algorithm (376)
QC tests required for new test strip lots
This configuration item specifies if a new test strip lot
can be used directly for patient testing, or whether a new
test strip lot must first pass QC tests successfully.
QC tests required for new test strip lots works in
conjunction with the QC algorithm configuration item.
Item ID
QC_NEEDED_FOR_NEW_STRIP_LOTS
Available values/format
Disabled (default)
[false]
A test strip lot can be used without passing a QC test.
Enabled
[true]
A test strip lot cannot be used without passing a QC
test.
Dependencies
Configuring QC tests required for new test strip lots
is only relevant when the QC algorithm configuration
item value is not Disabled.
Configuration
You configure the QC tests required for new test strip
lots configuration item using the following methods:
Data management system (e.g., cobas
®
infinity POC)
Barcode (generated via cobas
®
infinity edge)
10 Initial configuration